Implantation of the InVance®
male sling system is a minimally invasive procedure that typically
takes around 30-45 minutes. Following is a brief summary of the InVance
surgical procedure.
Caution: This information is for
educational purposes only. Refer to the surgical step procedure and
videos for more detailed instructions on this surgical procedure.
Step 1 Place
the patient in a modified lithotomy position. Insert a 14 Fr. Foley
catheter into the bladder to aid in urethral palpation. Palpate the
symphysis pubis to determine the incision site. Make a 3-4 cm midline
perineal incision.
Step 2 Carry out scissor
dissection laterally to expose the medial aspect of the descending pubic
rami on both sides. Expose the bone where the screws are to be placed.
Preserve the bulbospongiosus.
Step 3 Using
the InVance Inserter, place three AMS Screws with polypropylene sutures
in each rami. Place the proximal pair as high on the descending ramus as
possible, just beneath the pubic symphysis. Space the subsequent pairs 1
cm apart.
Step 4 Grasping the InVance
Inserter, press the screw firmly to the bone. Activate the Inserter
until the motor changes pitch, indicating that the screw is firmly
seated. Pull on the suture to ensure good capture.
Step 5 After
all screws are placed, cut the ends of each suture. The sling material
can be trimmed to ensure appropriate fit on the pubic bone. On one side,
pass sutures through the sling material with the aid of an 18 gauge
needle. Firmly tie the sling to the bone on that side.
Step 6 Using
a right angle clamp, pull the sling across the urethra to approximate
its secured position. Confirm tension using either a Cough Test (if
patient is awake) or Retrograde Perfusion Pressure (RPP) Test. Mark
the sling material with a sterile marker to facilitate suture
placement.
Step 7 Cough Test—(Only if patient is awake) Fill the bladder and ask the patient to give a strong cough to determine appropriate sling tension. RPP Test—Pull
back the catheter into the distal penile urethra and fill the balloon
with 1cc of saline. Connect the catheter to a saline infusion bag via
sterile tubing. Elevate the bag until infusion begins. Record the
retrograde leak point pressure in cm of water from the level of the
symphysis to the top of the fluid column (top of water level in saline
bag).
Step 8 Pass the remaining sutures
through the sling material. Tie the sling down completely, beginning
with the most anterior screw. Confirm appropriate sling tension and trim
all sutures. Close in a multi-layer closure.
Invance 8-Step Procedure Guide |